Pulse oximetry-based scoring may miss sleep apnea events in Black patients

A study published in the American Journal of Respiratory and Critical Care Medicine suggests that pulse oximetry may undercount breathing events in Black patients due to skin pigmentation interference. This bias could lead to inaccurate sleep apnea diagnoses and severity classifications.
Why it matters
Identifying racial biases in medical diagnostic tools is critical for ensuring equitable healthcare outcomes and improving the accuracy of clinical assessments.
edited by Sadie Harley , reviewed by Robert Egan
This article has been reviewed according to Science X's editorial process and policies . Editors have highlighted the following attributes while ensuring the content's credibility:
Add to Preferred Sources Credit: Unsplash/CC0 Public Domain Sleep studies often use a pulse oximeter to determine the severity of sleep apnea. But a new study by investigators from Mass General Brigham reports that pulse oximeters may undercount breathing events in Black patients, potentially leading to inaccurate severity classifications.
Researchers found that the proportion of airflow-defined breathing events meeting a standard oxygen-desaturation criterion was about 20% lower in Black than white participants. The findings were published in the American Journal of Respiratory and Critical Care Medicine .